Philips BiPAP A40 is designed to provide effective long-term therapy for patients suffering from conditions such as sleep apnea and other chronic respiratory challenges. The system is programmed to monitor the patient’s upper airway resistance and adjust EPAP delivery automatically to keep the airways open.
The device simultaneously monitors and adjusts
pressure support to maintain TTV (Target Tidal Volume). Set to Auto breath
rate, the device adjusts the backup breath rate based on the patient’s
spontaneous respiratory rate.
With patient comfort foremost in their minds,
designers have equipped the BiPAP A40 with the clinically proven Auto-TRAK
algorithm with an enhanced leak compensation technology to handle changes in
respiration.
Average Volume Assured Pressure Support (AVAPS) is
a feature available in the S, S/T, PC, and T modes. (In AVAPS-AE mode, the
AVAPS feature is always enabled.) AVAPS helps patients maintain a tidal volume
(VT) equal to or greater than the target tidal volume (Tidal Volume setting) by
automatically controlling the pressure support (PS) provided to the patient.
Bi-Flex Comfort Feature-If enabled, the device provides a comfort feature
called Bi-Flex in S mode only.
The Bi-Flex attribute adjusts therapy by inserting
a small amount of pressure relief during the latter stages of inspiration and
during active exhalation (the beginning part of exhalation). Ramp-The device is
equipped with an optional ramp feature.
The ramp feature is designed to offer lower
pressures when activated and then gradually increase to allow the patient to
fall asleep.
Airline Travel-The device is suitable for use on
airlines when it is operating from an AC or DC power source.
Ventilation modes : CPAP, S, S/T, PC, T, AVAPS-AE.
Target tidal volume : 200 – 1500 ML
Mode of Operation : Continuous
Operating Temperature :5° C to 35° C (41° F to 95°
F)
Storage Temparature :-20° C to 60° C (-4° F to 140°
F)
Relative Humidity :15 to 95% (non-condensing)
Atmospheric Pressure :101 kPa to 77 kPa
(approximately 0-2286 m/0-7500 ft)
AC Voltage Source : 100 to 240 VAC, 50/60 Hz, 1.2 A
DC Power Source : 12 VDC, 5.0 A (External Battery)
24 VDC, 4.2 A (Power Supply)
Data Management : EncorePro 2 and DirectView
software Compatible with oximetry module, Advanced detection of residual
respiratory events (Apnea/Hypopnea Index, Obstructed Airway Apnea, Clear Airway
Apnea, Hypopnea, Periodic Breathing, RERA, Large Leak and Snore)
Parameter Range Accuracy
IPAP 4 – 40 cm H2O ± 2.5 cm H2O*
EPAP 4 – 25 cm H2O ± 2.5 cm H2O*
CPAP 4 – 20 cm H2O ± 2.5 cm H2O*
Breath rate 0 to 40 BPM greater of ± 1 BPM or ±10% of setting
Inspiration time 0.5 to 3 seconds ± (10% of setting + 0.1 second)
Contact :
TriMedicare
1st Floor, 68/113A Jessore Road, Amarpally, Dum Dum, Kolkata – 700074
Email : trimedicare@aol.com
Call : (+91) 7003945473
(+91) 8777052658
(+91) 9883235935
(+91) 6291160936
0 Comments